MEASUREMENT FUNCTIONS
Parameters Measured: R, L, C, D, & Q.
Parameter Selection: Manual or automatic selection of R, L or C.
Measurement Modes: Series or parallel equivalent circuit.
Range Hold: Prevents autoranging when changing components.
Zero Function: Nulls out up to 100pF at test fixture.
Measurement Freq.: Selectable as 100Hz/120Hz, 1kHz, 10kHz.
Displayed Functions: R+Q, L+Q, C+D, C+R.
MEASURING RANGES
Resistance: 0.1mOhms to 990MOhms
Inductance: 0.001H to 9900H
Capacitance: 0.001pF to 99000uF
D: 0.001 to 999
Q: 0.001 to 999
| Accuracy
Limits: |
|
100Hz |
1kHz |
10kHz |
| Resistance |
0.1% ± 1
digit |
2Ohms-1MOhms |
2Ohms-500kOhms |
2Ohms-50kOhms |
|
0.5% ± 1
digit |
0.4Ohms-5MOhms |
0.4Ohms-2MOhms |
0.4Ohms-200kOhms |
|
2% ± 1
digit |
0.1Ohms-20MOhms |
0.1Ohms-10MOhms |
0.1Ohms-500kOhms |
| Inductance |
0.1% ± 1
digit |
4mH-500H |
0.4mH-50H |
40µH -5H |
|
0.5% ± 1
digit |
0.8mH-2500H |
80µH-250H |
8µH -25H |
|
2% ± 1
digit |
0.2mH-9900H |
20µH-1000H |
2µH -100H |
| Capacitance |
0.1% ± 1
digit |
10nF-1000µF |
1nF-100µF |
100pF-10µF |
|
0.5% ± 1
digit |
2nF-5000µF |
200pF-500µF |
20pF-50µF |
|
2% ± 1
digit |
500pF-20000µF |
50pF-2000µF |
5pF-200µF |
(R accuracies apply for Q<0.1. L
accuracies apply for Q>10. C accuracies apply for D<0.1 and after
Null).
LIMITS COMPARATOR
Limits Set-up: Multiple Upper and Lower limits can be set from
keyboard or from RS-232 interface
Binning: Up to 8 Pass bins can be defined plus Fail on minor parameter and
general Fail.
COMPONENT CONNECTION
Component Connection: 4-terminal connection via internal fixture for
both radial and axial components.
External 4-terminal connection via adaptor.
Internal Fixture: Sprung jaws for vertical lead insertion. Plug-in sliding
pillars for horizontal lead insertion.
Bias Voltage: Switchable 2V polarising voltage. External bias up to 50V.
DISPLAY
Display Type: Dual 5-digit 0.56” LEDs.
Annunciators: LED annunciators for all functions and measurement units.
Displayed Functions: R+Q, L+Q, C+D, C+R, PASS/FAIL, Bin No.
DIGITAL INTERFACE
Interface Type: RS-232 via 9 pin D connector, 9600 baud.
Interface Function: Full command and readback.
GENERAL
Keyboard: Full numeric keyboard.
Non-volatile memory: Up to 9 complete set-ups.
Input Voltage: 230V or 115V ±14%, 50/60Hz, internally adjustable, 25VA
max. Installation Category ll.
Temperature Range: +5oC to 40oC operating 20 - 80% RH, -40oC to 70oC
storage.
Safety: Complies with EN61010-1.
EMC: Complies with EN55081-1 and EN55082-1.
Size: 360 x 240 x 95 mm including feet.
Weight: 2.9kg approx.
OPTIONS
LCR-SMD: Surface Mount Tweezers (4 terminal). Plugs directly into the
standard component fixture on the LCR400.
LCR-BNC: Plugs into the standard component fixture to convert it into a set of
four BNC connectors.
Supplied complete with four BNC to BNC leads.
LCR-K-Clip: High quality Kelvin Clip test lead set terminating in BNC plugs
(requires LCR-BNC).
LCR-PCLink: Windows based software which logs results (values and binning) and