| 1906 -
Description
and Specifications |
|
Greater resolution and accuracy
The 1906 is a true 5.5 digit meter with a scale length of ± 210,000
counts. This gives it not just superior resolution relative to 4.5 digit
meters, but much higher effective accuracy as well.
Greater sensitivity and higher input impedance
With a maximum resolution of 1mV, 1mW and 1nA, the 1906 is ten times more
sensitive than a 4.5 digit meter.
This extra senstivity enables accurate measurements to be made in areas
previously impossible such as thermocouple junctions, switch contact
resistance or capacitor leakages.
4 terminal Ohms for improved stability
Resistance measurements can be made in either 2 terminal or 4 terminal
mode. High impedance sensing ensures that the meter switches automatically
between modes. For low resistance measurements, 4 terminal mode offers
much higher accuracy and repeatability.
True RMS ac functions
The 1906 provides True RMS ac response which gives accurate measurements
regardless of the waveform shape. The wide bandwidth voltage attenuator
provides high accuracy within the audio band and gives extended response
to avoid distortion when measuring switching waveforms.
Full bus control via RS-232 or GPIB
The 1906 is designed for complete bus control. Every function of the
instrument can be controlled using either the RS-232 interface or the GPIB
interface (optional).
Wide range of Smart functions
The 1906 offers computing and datalogging functions as standard: dBs, Ax+b,
Limits, % deviation, Min-Max, and a 100 reading Data-logger.
Auto or manual ranging
The 1906 offers fully automatic ranging on all functions including
current. Alternatively any range can be selected and held manually.
Closed case calibration
The 1906 stores its calibration constants in permanent memory.
Consequently it can be recalibrated without ever opening the case.
Storage of front-panel set-ups
The 1906 can store up to six front panel set-ups in non-volatile memory.
In addition to the range and function each set-up defines the active
computing or logging functions along with their parameters.
Single key convenience functions for ease of use
Press the “null” key and the present displayed reading is stored and
subtracted from all future readings. This powerful feature has a host of
uses such as removing test lead resistance, observing deviation levels, or
measuring relative to alternative voltage points using a fixed ground lead
connection.
Full protection for the instrument and the user
The 1906 incorporates extensive protection against damage from accidental
overloads on all ranges including 10 Amps.
In addition the 1906 meets the stringent safety requirements of IEC348 and
IEC1010-1 for measurements up to 1kV.
|
|
Touch and Hold
The touch and hold function is invaluable when taking measurements in
awkward situations where it may be difficult to see the multimeter
display. With touch and hold selected, the display is automatically frozen
every time a stable reading is achieved. A further press of the T/HOLD key
un-freezes the display.
Digital filtering for better noise performance
The 1906 uses non-linear digital filtering which is highly effective in
eliminating unwanted noise, but responds rapidly to a change in the
reading. When working with very noisy or unstable measurements, the
characteristics of the digital filter can be adjusted to suit the
application.
Linear scaling with offset (Ax+b)
This function enables a reading to be multiplied by a scale factor (A) and
for an offset (b) to be added or subtracted. This means, for example, that
the electrical output of a transducer can be scaled to give a direct
readout of the physical parameter e.g. temperature or weight. It is
particularly useful with 4-20mA current loops.
Limits comparison
This function enables the reading to be tested against high and low limits
set by the user. The display then shows a code of H (high), L (low) or P
(pass) at the end of the display.
Percentage deviation (D%)
This function enables the meter to display the percentage amount by which
the reading differs from a nominal value chosen by the user. It is
invaluable for measuring tolerance or stability.
Min-Max storage
The Min-Max function stores the highest and lowest values of a set of
readings. This facility has many uses such as checking for power supply
glitches, recording peak temperature excursions and avoiding “missed”
readings when using the Logger function.
Logarithmic measurements (dB)
This function allows measurements to be displayed in the form 20log10X.
Built-in zero reference scaling enables voltages to be displayed in dBV or
dBm relative to any required impedance. Using the null key allows gains
and losses to be measured directly.
Automatic data logging
The 1906 can store up to 100 readings at any required time interval from 1
reading per second up to 1 reading every 3 hours.
A simple recall sequence allows the readings to be scrolled onto the
display whenever required. Alternatively results can be downloaded using
the RS232 or GPIB interfaces. As a result, tedious time related
measurement sequences can be handled automatically without disrupting your
work.
Manual storage of readings is also available eliminating the need for
paper and pencil when making a series of measurements. Storage can also be
triggered from the RS-232 or GPIB interfaces.
|
|
|
|
|
|
ACCURACY
Accuracies apply
for 1 year 18oC to 26oC.
Temp. coefficient outside these limits is <0×1 x quoted range
accuracy per oC. Warm-up time to
rated accuracy is 1 hour.
DC
VOLTS
|
Range
|
Accuracy(±)
|
Input
Impedance
|
Max
Input
|
|
200mV
|
0×017% + 3d
|
10M or >1G
(Hi Z mode)
|
1kV DC or AC
Peak any range
|
|
2V
|
0×012% + 3d
|
|
|
20V
|
0×019% + 3d
|
10M
|
|
200V
|
0×019% + 3d
|
|
|
1000V
|
0×019% + 3d
|
|
RESISTANCE
(OHMS)
|
Range
|
Accuracy(±)
|
Max. Measuring
Current
|
Max.
Input
|
|
200
|
0×025% + 4d
|
1×5mA
|
300V DC or
RMS continuous any Ohms input any range
|
|
2k
|
0×019% + 3d
|
1×5mA
|
|
20k
|
0×019% + 3d
|
0×4mA
|
|
200k
|
0×019% + 3d
|
40mA
|
|
2M
|
0×022% + 3d
|
4mA
|
|
20M
|
0×07% + 3d
|
400nA
|
Max open
circuit voltage 3×5V. Diode Test on 2k range.
Audible continuity check available on all ranges.
AC
VOLTS (True RMS)
|
Range
|
Accuracy(±)
|
|
|
45Hz - 5kHz
|
5kHz - 10kHz
|
10kHz-20kHz
|
20kHz-50kHz
|
|
200mV
|
0×2% + 150d
|
0×2% + 250d
|
3% + 500d
|
-
|
|
2V
|
0×2% + 150d
|
0×2% + 250d
|
0×5% + 300d
|
3% + 1500d
|
|
20V
|
0×2% + 150d
|
0×2% + 250d
|
0×5% + 300d
|
2% + 1000d
|
|
200V
|
0×2% + 150d
|
0×2% + 250d
|
0×5% + 500d
|
-
|
|
750V
|
0×2% + 150d
|
-
|
-
|
-
|
Additional
error at crest factor=3 typically 0×7%, Input impedance 1M / <100 pF any
range.
Max input 750V rms, 1kV peak, any range.
DC
AMPS
|
Range
|
|
Accuracy(±)
|
Voltage
Burden
|
Max.
Input
|
|
200mA
|
|
0×08%
+ 12d
|
300 mV max
|
1A, 300V fuse
protected
|
|
2mA
|
|
0×08%
+ 12d
|
|
|
20mA
|
|
0×08%
+ 12d
|
|
|
200mA
|
|
0×08%
+ 12d
|
|
|
10A
|
<2000mA
|
0×08% + 12d
|
650mV max
|
10A, 300V
fuse protected
|
|
|
>2000mA
|
0×12% + 12d
|
|
AC
AMPS (True RMS)
|
Range
|
Accuracy(±)
45 Hz - 1 kHz
|
Voltage
Burden
|
Max
Input
|
|
200mA
|
|
0×37%
+ 100d
|
300mV max
|
1A, 300V fuse
protected
|
|
2mA
|
|
0×37%
+ 100d
|
|
|
20mA
|
|
0×37%
+ 100d
|
|
|
200mA
|
|
0×37%
+ 100d
|
|
|
10A
|
<2000mA
|
0×37% + 100d
|
650mV max
|
10A, 300V
fuse protected
|
|
|
>2000mA
|
0×4% + 100d
|
|
Additional
error at crest factor=3 typically 0×7%.
FURTHER SPECIFICATIONS
INPUT CHARACTERISTICS
Input Current: < 100 pA.
DC NMR: > 60dB at 50/60Hz.
1k Unbalanced CMR: > 120dB at DC/50Hz/60Hz, DC ranges; > 60dB at
DC/50Hz/60Hz, AC ranges.
DISPLAY
Display Type: 13mm LED, 8 digit.
Scale Length: Selectable 5˝ digit or 4˝ digit.
Reading Rate: 3 readings/sec (5˝ digit). 10 readings/sec (4˝ digit).
Overrange: Displays OL if input too great for range.
Overflow: Displays OFLO if calculated result overflows display.
Annunciators: LEDs for range, function and program modes.
COMPUTING FUNCTIONS
Null: Operates over full range; values can be stored for every function.
Digital Filter: 10 options, including optimised default values for each
range and function.
Touch & Hold: Reading is held when stable.
% DEV: Displays % deviation from entered reference value.
dB: Displays measurement in dB relative to 1V, 1mA, 1kW or user entered
value, or in dBm.
Ax+B: Linear scaling of results, with offset.
Limits: Reading displayed with H, L, or P (pass) with respect to
user-defined high and low limits.
Min/Max: Minimum and maximum reading stored.
Data Logger: Manual or automatic storage of 100 measurements, Storage
interval 1s to 9999s.
INTERFACES
RS232: Baud rates 300, 1200 or 9600. Complies fully with the
ARC
(Addressable RS232 Chain) interface standard. Address selectable from the
front panel.
GPIB (Optional): Fully complies with IEEE-488.1 and IEEE-488.2
POWER REQUIREMENTS
AC Input: 110/120 Volts AC nominal 50/60Hz or 220/240 Volts AC nominal
50/60Hz by rear panel adjustment. The instrument will operate safely and
meet specification within normal AC supply variations, viz. 99-132 Volts
AC and 198-264 Volts AC respectively.
Consumption: 25VA max.
GENERAL
Operating Range: +5oC to +40oC, 20% to 80% RH
Storage Range: -40oC to +60oC
Size: 260(W) x 88(H) x 235(D)mm, excl. handle and feet
Weight: 2×2kg.
|
|
Thurlby Thandar Instruments Ltd. operates a policy of continuous development
and reserves the right to alter specifications
without prior notice.
|
 |
|