The Fluke Handheld Test Tools range is available from TTi in the UK

  Fluke VR101S Voltage Event Recorder System - Specifications

   Set up, plug in, download, and analyse
  • Generates professional power quality reports
  • Monitors receptacles, the source of power for many critical loads
  • VR101R includes everything Applications need to get started including EventView™ software for Windows and an interface cable
  • Stores up to 4000 events
  • Recorder with a PC and EventView Software
  • Does not need to remain connected to the PC to capture events
  • Additional VR101 Event Recorders can be purchased separately for monitoring multiple locations at once
  • Capture limits are user-selectable by loading them into the VR101 Event
EQUIPMENT RATINGS
VR101/xx3
Nominal Supply 120V
Operating Range 70 to 140V
Operating Frequency 50/60 Hz
Power 2W
VR101/xx1, /xx4, /xx5, and /xx7
Nominal Supply 240V
Operating Range 140 to 270V
Operating Frequency 50/60 Hz
Power 3W
Memory Size 32kB, 4000 events
Storage Method First-in First-out, or Fill-then-stop
Sampling Method Continuous (half cycle integrated)
Alarm Type Optical; red LED slowly blinks when there are events stored in memory.
Polarity Detection Within 8 s after plugging in.
Polarity Indication Red LED lights steadily during 8 to 16 seconds if polarity is OK.
Red LED quickly blinks during 8 to 16 seconds if improper polarity.
Size 85 mm X 68 mm X 35 mm
Weight 120g (4 oz)
Operating Limits -40 to 160°F (-40 to 70°C) (excluding Optical Interface Cable)
0 to 95% RH (non-condensing)
15,000 feet (4.57 km) altitude.
Battery 3.6V Lithium (non-replaceable)
Expected Battery Life 7 years (provided that maximum battery care is taken)
Fuse 1/16A Fast-Blow (non-replaceable)
SAG, SWELL AND OUTAGE MEASUREMENT
Hot to Neutral
VR101/xx3
Range 0 to 200 Vrms
Accuracy ±2 Vrms*
Resolution 1 Vrms
VR101/xx1, /xx4, /xx5, and /xx7
Range 0 to 270 Vrms
Accuracy ±4 Vrms**
Resolution 2 Vrms
Neutral to Ground
VR101/xx3
Range 3 to 150 Vrms
Accuracy ±2 Vrms
Resolution 1 Vrms
VR101/xx1, /xx4, /xx5, and /xx7
Range 3 to 150 Vrms
Accuracy ±2 Vrms
Resolution 1 Vrms
*sags less than 70V for > 1s will be reported as 0V
**sags less than 140 V for > 1s will be reported as 0V
TRANSIENT MEASUREMENT
Range
Hot to Neutral 100 to 2500V peak
Neutral to Ground 50 to 2500V peak
Accuracy ± 10% of reading + 10V
Resolution 10V
Width Detection 1 µs minimum
Phase Angle
Range 20° to 180°, 200° to 360°
Accuracy ±1°
Resolution 1°
Frequency Measurement
Range 45 to 65 Hz
Accuracy ±0.1 Hz (3 cycles minimum)
Resolution 0.1 Hz
TIME MEASUREMENT
Events <1 s(econd)
Accuracy
Hot to Neutral ±0.5 cycle*
Neutral to Ground ±1 cycle**
Resolution
Hot to Neutral 0.5 cycle
Neutral to Ground 1 cycle
Clock Accuracy ±2 s/day
Timer Counter Capacity 4.25 year (24 bit counter, resolution 8 seconds)
*Outages are reported with 1/2 cycle resolution until the VER goes into low power mode.  A VER will go into the low power mode after approximately one second of an outage condition.  In low power mode, the resolution becomes 8 seconds.  Swell, sag, and frequency events have 1/2 cycle resolution up to approximately 10 minutes after which the resolution becomes 8 seconds.  **Neutral to Ground swells have 1 cycle resolution up to approximately 10 minutes after which the resolution becomes 8 seconds.
APPROVALS
Safety This device is safety approved by CSA and CSA (NRTL/C) for measurements on 300V CAT III, Pollution Degree 2, double insulation, per: UL3111-1 (1994)  IEC1010-1 (1990)
CSA C22.2 No. 1010.1 (1992)  EN61010-1 (1993)
Environmental MIL-T-28800E, Type III, Class 3
EMC FCC rules part 15.  89/336/EEC  Emission: EN50081.1
Immunity: EN50082.2  (IEC1000-4-2, -3, -4, -5)

Thurlby Thandar Instruments Ltd.   Glebe Road, Huntingdon, Cambridgeshire PE29 7DR  U.K.
Tel: 01480 412451  Fax: 01480 450409  Email: sales@tti.co.uk   