Technical
Specifications
|
| Display |
|
| Resolution |
|
| V DC |
| Ranges: |
100 mV to 1000 V |
| Max. Resolution: |
100 nV |
| Accuracy: |
8845A: 0.0035 +
0.0005 |
|
8846A: 0.0024 +
0.0005 | |
| V AC |
| Ranges: |
8845A: 100 mV to 750
V |
|
8846A: 100 mV to 1000
V |
| Max. Resolution: |
100 nV |
| Accuracy: |
8845A: 0.06 + 0.03 |
|
8846A: 0.06 + 0.03 |
| Frequency: |
3 Hz to 300
KHz | |
| Resistance |
| 2x4 Wire: |
Yes |
| Ranges: |
8845A: 100 Ω to 100
MΩ |
|
8846A: 10 Ω to 1 GΩ |
| Max. Resolution: |
8845A: 100 μΩ |
|
8846A: 10 μΩ |
| Accuracy: |
8845A: 0.010 + 0.001 |
|
8846A: 0.010 +
0.001 | |
| A DC |
| Ranges: |
100 μA to 10 A |
| Max. Resolution: |
100 pA |
| Accuracy: |
0.050 +
0.005 | |
| A AC |
| Ranges: |
8845A: 10 mA to 10 A |
|
8846A: 100 mA to 10 A |
| Max. Resolution: |
8845A: 10 nA |
|
8846A: 100 pA |
| Accuracy: |
0.10 + 0.04 |
| Frequency: |
3 Hz to 10
kHz | |
| Freq/Period |
| Ranges: |
8845A: 3 Hz to 300
kHz |
|
8846A: 3 Hz to 1 MHz |
| Max. Resolution: |
1 μHz |
| Accuracy: |
0.01
% | |
| Continuity/ Diode
Test |
|
| Capacitance |
| Ranges: |
8846A: 1 nF to 0.1 F |
| Max. Resolution: |
8846A: 1 pf |
| Accuracy: |
8846A: 1
% | |
| Temperature |
| Type: |
8846A: Platinum RTD |
| Ranges: |
8846A: -200 °C to +600
° |
| Max. Resolution: |
8846A: 0.01 ° |
| Accuracy: |
8846A: 0.06
° | |
| Math Functions |
|
NULL, Min/Max/Average, Std
Dev |
| dB/dBm: |
Yes | |
| Advanced Functions |
| Statistics/Histogram: |
Yes |
| Trendplot: |
Yes |
| Limit Test: |
Yes | |
| Input Output |
| USB Memory: |
8846A: USB memory drive
port |
| Real Time Clock: |
8846A: Yes |
| Interfaces: |
RS 232, IEE-488.2,
Ethernet, USB (with optional adaptor) |
| Programming
Languages/Modes: |
8846A: SCPI (IEEE-488.2),
Agilent 34401A, Fluke
45 | |
|
Weight |
|
| Dimensions |
|
88 mm x 217 mm x 297 mm
(3.46 in x 8.56 in x 11.7
in)
| |
| Safety |
|